Colorimetric capnography is a qualitative measurement method that detects the presence of carbon dioxide (CO2, a relatively acidic gas) in a given gaseous environment. From a medical perspective, the method is usually applied by exposing litmus paper/film to an environment containing a patient's airway gases (i.e. placing it into their breathing circuit/airway circuit), where it will then change color depending on the amount of exhaled CO2 within the circuit. Although there are multiple uses, the colorimetric capnography method is most frequently used to quickly confirm that an advanced airway device such as an endotracheal tube (ETT) or nasotracheal tube (NTT) has been placed in the desired location. Correct placement is evidenced by sufficient color change of the litmus paper while exposed to the airway gases which, if the device is properly placed, will contain relatively high amounts of exhaled CO2. Conversely, lack of detection of CO2 suggests many possible issues, including improper placement of the advanced airway device, most commonly into the esophagus.[1] There are no contraindications to capnography.[2]

Equipment

Colorimetric capnography devices do not require electricity, change color reversibly (breath-by-breath), and add a small amount of dead space. A variety of devices are available from multiple manufacturers. Specific devices on the market include the Covidien Nellcor EasyCap or PediCap (manufactured by Medtronic), and the StatCO2, Mini StatCO2, and Neo-StatCO2 (made by Mercury Medical).

Metacresol purple is an example of a pH-dependent dye commonly used in such devices. The colour change of the EasyCap is from purple to yellow/gold, leading to the phrase 'good as gold' or 'gold is good', as an aide-mémoire for successful tracheal intubation.

Medical applications

Airway device placement

Colorimetric capnography is most frequently used in medical contexts to determine whether exhaled CO2 is present in a patient's airway circuit after placement of an advanced airway device (e.g. ETT or NTT), but the meaningful significance of CO2 detection is in reality much more broad. Detection of sufficient CO2 suggests that a patient is metabolically active and capable of eliminating CO2, while lack of color change or insufficient color change suggests an issue at one or more of many possible levels including issues with cellular metabolism, airway device integrity and placement, circuit integrity, mucous plugging, cardiopulmonary function, device malfunction, and more.[1]

Nasogastric tube placement

Inadvertent placement of a nasogastric tube (NGT) into the airway rather than the stomach can lead to complications such as pneumothorax and pneumonia. The use of colorimetric capnography to detect proper placement of a nasogastric tube (NGT) has shown promise, especially in mechanically ventilated patients. With this method, rather than exposing the litmus paper to the patient's airway gases via connection with the airway circuit, the litmus paper is exposed to the gases transmitted via the nasogastric tube. Put simply, just as this method can detect the desired presence of significant CO2 in an airway, it can also detect the desired lack of CO2 in a properly placed nasogastric tube.[3][4] Further research on this particular use of colorimetric capnography and the widespread incorporation of adapters will likely be required if this utilization is to become more commonplace.[4]

Patient monitoring

Data inputs from multiple sources (e.g. blood pressure cuff (sphygmomanometer), pulse oximeter, thermometer, etc.) are used to characterize a patient's vital condition, and the more meaningful data that is available, the more accurate and precise a clinician can be in addressing a patient's health status. Capnography as a whole represents a significantly useful data point in monitoring a patient's respiratory and metabolic status in situations including but not limited to cardiac arrest, metabolic acidemia, mechanical ventilation, and procedural sedation.[5]

Comparison to quantitative capnography

One disadvantage of qualitative capnography methods such as colorimetry is that they do not produce a direct numeric or waveform readout as can a quantitative method such as infrared capnography. The colorimetric method rather presents CO2 simply as a color most commonly on the spectrum of purple (lower CO2) to gold (higher CO2) and leaves the clinician to interpret anything beyond this single returned value. Despite being a more rudimentary method of capnography as compared to quantitative methods such as infrared capnography, colorimetric capnography has proven to remain beneficial in multiple contexts in modern medicine as mentioned in the above section.[5] Evidence also suggests that colorimetric capnography is just as effective as infrared capnography at determining correct tracheal airway device placement, but may fall short of infrared capnography when detecting inadvertent esophageal airway device placement.[6]

Naming controversy

There has been debate related to the term 'colorimetric capnography,' especially with the growing employment of this method during the COVID-19 pandemic due to an increased volume of patients requiring mechanical ventilation. The use of the term 'capnography' technically implies some type of quantitative result or readout which, as described above, is not produced by the colorimetric method. Descriptors such as "carbon dioxide colorimetry," "colorimetric carbon dioxide detection," and "chemical colorimetric analysis" have been suggested as potentially more accurate replacements.[7]

See also

References

  1. 1 2 Nowicki, Thomas; Jamal, Zohaib; London, Shawn (2023), "Carbon Dioxide Detector", StatPearls, Treasure Island (FL): StatPearls Publishing, PMID 29489155, retrieved 2023-12-04
  2. Butterworth, John F. IV; Mackey, David C.; Wasnick, John D. (2022). Morgan and Mikhail's Clinical Anesthesiology (7th ed.). McGraw Hill LANGE. pp. 115–116. ISBN 978-1-260-47379-7.
  3. Chau, Janita Pak Chun; Liu, Xu; Choi, Kai Chow; Lo, Suzanne Hoi Shan; Lam, Simon Kwun Yu; Chan, Kam Ming; Zhao, Jie; Thompson, David R. (November 2021). "Diagnostic accuracy of end-tidal carbon dioxide detection in determining correct placement of nasogastric tube: An updated systematic review with meta-analysis". International Journal of Nursing Studies. 123: 104071. doi:10.1016/j.ijnurstu.2021.104071. ISSN 1873-491X. PMID 34520886. S2CID 237514424.
  4. 1 2 Bennetzen, Linda Vad; Håkonsen, Sasja Jul; Svenningsen, Helle; Larsen, Palle (January 2015). "Diagnostic accuracy of methods used to verify nasogastric tube position in mechanically ventilated adult patients: a systematic review". JBI Database of Systematic Reviews and Implementation Reports. 13 (1): 188–223. doi:10.11124/jbisrir-2015-1609. ISSN 2202-4433. PMID 26447016. S2CID 3957809.
  5. 1 2 Long, Brit; Koyfman, Alex; Vivirito, Michael A. (December 2017). "Capnography in the Emergency Department: A Review of Uses, Waveforms, and Limitations". The Journal of Emergency Medicine. 53 (6): 829–842. doi:10.1016/j.jemermed.2017.08.026. ISSN 0736-4679. PMID 28993038.
  6. Hogg, K.; Teece, S. (May 2003). "Towards evidence based emergency medicine: best BETs from the Manchester Royal Infirmary. Colourimetric CO(2) detector compared with capnography for confirming ET tube placement". Emergency Medicine Journal: EMJ. 20 (3): 265–266. doi:10.1136/emj.20.3.265. ISSN 1472-0213. PMC 1726107. PMID 12748149.
  7. Canelli, Robert; Ortega, Rafael (2021). "Colorimetric capnography: a misnomer worth correcting". Journal of Clinical Monitoring and Computing. 35 (4): 951. doi:10.1007/s10877-021-00665-5. ISSN 1387-1307. PMC 7870361. PMID 33558982.
This article is issued from Wikipedia. The text is licensed under Creative Commons - Attribution - Sharealike. Additional terms may apply for the media files.